Specifications:
SpO2:
- Range: 0-100%
- Resolution: 1%
- Accuracy:(70-100%) +/-2% Adult
- Method: Dual wavelength LED
Heart (Pulse) RATE:
- Range: 25-254bpm
- Resolution: 1bpm
- Accuracy: 2bpm
Display:
- Type: Backlit LCD
- Parameters SpO2; PR
- Status: Standby; Communication; Download
- Information: Patient ID; Indicator of data storage; Low power; Pleth. Bar; Error; Standby
Mechanical:
- Weight: 130g without batteries and sensors
- Size: 65 x 140 x 32mm(WxHxD)
Battery:
- Type: 4 AA Alkaline batteries or 4 AA rechargeable batteries
- Life: Approximately 15 hours for normal measurement status
Classification:
- Type of Protection: Internally powered equipment
- Degree of Protection: Before (BF)
- Safety: IEC 60601-1, EN 865, ISO 14971
Storage:
- Patient ID: Up to 100
- Data records: Up to 200
|