The Capnocheck® Sleep Capnograph/Oximeter is specially designed for sleep screenings and lab-based sleep studies. Its 4-channel analog output capability lets you select ETC02 numeric value, ETC02 waveform, inspired C02, respiratory rate, heart rate, as well as Sp02 numeric and waveform displays. This next generation system utilizes Serial Autocorrelation (SAC) digital pulse oximetry technology for reliable SpO2 measurements, and the breath by breath CO2 detection and display is ideal where fast response to patient change is needed.
Features
- Quiet System minimizes monitor noise
- Patented technology reliably measures Sp02
- Breath-by-breath CO2 detection allows rapid response to patient changes
- Disappearing display and no audible alarms or pulse tones minimize patient distraction
- 4-channel analog output capability for comprehensive CO2 and Sp02 monitoring in one unit
- Analog signal carrying capacity offers versatile room and lab set-up flexibility
- Perform non-lab or preliminary steep studies via PC analysis software
- Incorporates economical sidestream technology
Specifications
- Display: Vacuum Fluorescent
- Indicators: Alarm, Alert, Charge, Silenced
- Flow Rate: 150 ml/min, +/- 20
- CO2:
- Measurement: Non-Dispersive IR absorption
- Measurement Range: 0-10% C02
- Units: User-selectable: mmHg, Vol. % kPa
- Display Range: 0-100 mmHg or 0-10.0% C02 or 0-13.3 kPa
- Accuracy: +/-2 mmHg; +/-0.3%; +/-0.3 kPa, or 4% of reading, whichever is greater (0-10%CO2)
- Resolution: 1 mmHg; 0.1%; 0.1 kPa
- Calibration: 2 point Cal
- N20
- Compensation: User-selectable 40% (default=OFF)
- Response Time: 375 msec to 90% of value
- Averaging: 4-breath average, or breath-to-breath selectable
- Respiration Rate:
- Range: 0-150 bpm
- Averaging: 4 breath average
- Accuracy: +/- 1 bpm
- SpO2:
- Range: 0-100% Functional SpO2
- Accuracy: +/-2% at 70-100%: <70% unspecified
- Averaging: 2, 4, 8, or 16-beat averaging (default: 2-beat)
- Pulse Rate:
- Range: 30-254 bpm
- Accuracy: Greater of +/-2 bpm or +/- 2%
- Averaging: 8
- Pulse Tone: Pitch corresponds Sp02 value Volume adjustable or OFF
- Power Requirements:
- AC Charger Input 115V-, 60 Hz
- Battery: Nickel Cadmium battery, 6 VDC, internal rechargeable, not user replaceable, fully charged continuous- use life 2 hours approximate, maximum full charge time 4 hours
- Analog Output:
- Any four of: C02 waveform, Plethysmogram, End Tidal C02, Resiration Rate, Inspired C02, Sp02, Pulse Rate, 0V or 1V Cal Signal
- Range: 0-1 VDC full scale
- Serial Output:
- RS232C: tabular trends, patient data
- For Sleep Study: 10 hrs. at 4 sec., 31 hrs. at 12 sec.
- Dimensions: 3.5″Hx10.0″Wx5.5″D (88.9 mm H x 254mm W x 139.7mm D)